| Software Name | <Name> |
| Software Version | <x.x.x> |
| Software Location | <url> |
| Processes | <processes in which this tool is used> |
Describe intended use and usage context (e.g. automation, testing, control, altering). Include technical and usage requirements that the system shall fulfil.
Rate these aspects with yes (y) or no (n). If any of these aspects is rated as yes, the system is quality relevant and should be validated.
| Criterion | Y/N |
|---|---|
| Is the system used in one or more processes that steer the QMS? | |
| Could the conformity of the organization’s medical devices be affected if the system does not work according to its specifications? | |
| Could risks arise for patients, users, third parties or the organization if the system does not work according to its specifications? | |
| Does the software generate or manage data / records that are relevant to the QMS or medical device approval by authorities? | |
| Is the software used to generate electronic signatures on documents or records required by the QMS and/or state authorities? |
- Infrastructure software (e.g. operating systems, databases, office applications, antivirus, network management software) (GAMP category 1)
- Non-configurable software (GAMP category 3)
- Configurable software (GAMP category 4)
- Custom (self-developed) software (GAMP category 5)
List of Risks:
- <list of risks>
List of Risk Mitigation Measures (if necessary):
- <list possible risk mitigation measures>
Refer to section 10 for descriptions of the criticality classifications. If a software is not highly critical and widely adopted / commonly used, it can be continuously re-validated during use.
- Low (review upon major changes)
- Moderate (review every year)
- High (review every 6 months)
Provide a clear reason for the chosen criticality
| Role | Name | Task(s) |
|---|---|---|
- Software tool accessed with <Windows 10 20H2 on Google Chrome 88.0.4324.150>
- Reference User Manual
- Run software system on sample data
The software is approved for use if it is validated successfully and works as expected.
| ID | Expected | Result | Pass? |
|---|---|---|---|
| U1 | e.g. "A radiologist can log in with his/her email and password." | "Login with correct email and password grants access to the annotation tool." | yes |
| ID | Expected | Result | Pass? |
|---|---|---|---|
| T1 | e.g. "Execute correctly in the specified runtime (Google Chrome)." | "The application runs correctly in Google Chrome." | yes |
| Type | Total | Pass | Fail |
|---|---|---|---|
| Usage Requirements | 1 | 1 | 0 |
| Technical Requirements | 1 | 1 | 0 |
Approving the software for use is recommended due to the acceptance criteria being fulfilled completely.
You can optionally insert screenshots for proof of validation. Strictly speaking, this is not a hard requirement by the standards but it's nice to show when you're being audited.
| U1 | <insert screenshot> |
| T1 | <insert screenshot> |
| Date of Approval | Name of Approver |
|---|---|
| <date> | <name> |
| Date | Name | Activity |
|---|---|---|
| <Initial Approval> |
Criticality High
- A software failure can lead to physical harm requiring medical intervention
- Software controls parameters or data that are essential during product release
- Software manages data relevant for clinical evaluation or product approval
- Software manages data from which conclusions about incident messages or recall actions are drawn
Criticality Moderate
- A software failure can lead to physical damage requiring medical intervention
- Software administers documents whose loss endangers the certification
- Software controls intermediate results in the product realization, which are revealed in later steps by other processes
Criticality Low
- Software manages documents that play a role in the QM system, and whose loss would lead to an audit variance
Template Copyright openregulatory.com. See template license.
Please don't remove this notice even if you've modified contents of this template.